Control logic for a heart-lung machine's gas blender, written to IEC 62304 class C
We were part of the team that developed the electronic gas blender for the successor to the flagship heart-lung machine from LivaNova. We wrote the blender's control logic and its microcontroller drivers to IEC 62304, software safety class C.
- Air valve
- 2.03L/min
- Oxygen valve
- 1.97L/min
- To the oxygenator
- 4.0L/min· 60 % oxygen
At 60 % oxygen and 4.0 L/min the air valve lets through 2.03 L/min and the oxygen valve 1.97 L/min.
- Client
- LivaNova
- Industry
- Medical devices, cardiac surgery
- Our scope
- Control logic and microcontroller drivers of the electronic gas blender
- Standard and stack
- StandardIEC 62304
- Safety classC
- TargetMicrocontroller
Challenge
In class C, a failure of the software can lead to death or serious injury
During heart surgery a heart-lung machine does the work of the heart and the lungs. Its gas blender mixes air and oxygen for the oxygenator, the part of the machine where the blood takes up oxygen and gives off carbon dioxide.
IEC 62304 is the standard for medical device software. It sorts software into three safety classes, and the class decides how much of the standard's process applies. Class C is the highest, and all of it applies.
IEC 62304 · the three software safety classes
- A
No injury or damage to health is possible.
- B
Injury is possible, but not serious injury.
- C
Death or serious injury is possible.
The class of the gas blender's software
The worse the possible harm, the higher the class
What we did
We wrote the control logic and the drivers
The control logic decides what the blender does. The drivers connect it to the inputs and outputs of the microcontroller.
Under class C, each unit of code gets a written design before it is programmed and its own test afterwards. Every requirement has to be traceable to the test that proves it.
Control logic
Written by us
Decides what the device does.
Drivers
Written by us
Connect the logic to the inputs and outputs of the chip.
Microcontroller
Hardware, not ours
The chip the software runs on.

Oleksii Kvasnikov · Founder and lead engineer
Medical device software to IEC 62304